How to interface Human Humastar with LabXpert Automation

Ernest Okot

Last Update há 10 meses

LabXpert Automation connects with the Human Humastar by reading the contents of the folder `C:\ProgramData\HI\Human\LIS\ASTM\Output Worlist` for ASTM files. Therefore, the application must be installed only on the Humastar computer, otherwise it will not be able to access the folder.


This guide assumes you have installed the LabXpert Automation application and added the Human Humastar. See How to interface lab equipment with LabXpert Automation for more information.


  1. On the detail page of your Humastar in the application, click the setting icon on the top right
  2. In the popup, click the folder input and browse to a folder  `C:\ProgramData\HI\Human\LIS\ASTM\Output Worlist`



Approval of Worklist

Result worklist can be approved when the test results are all in test status "completed"

Worklist can't accept to approve/ send results to automation result tab if there are ongoing test run "Executing/scheduled"

Upon  clicking  approve button, a new tab will pop out and choose by clicking

(1) "approve only tests already completed"

This choice only works when they are no ongoing runs or scheduled run

it allows several approval of completed tests.

Upon executing /scheduling new runs, worksheet approved status changes to approve.


(2) "Definitively approve worksheet"

This choice make approve button default without need to click it again.

This make allows all tests under this worksheet to be automatically approved. Upon executing /scheduling new runs, worksheet approved status remain.


Note: After executing these choice, new scheduled runs can be added/executed/scheduled onto these worksheet